Agilent Technologies N9340A Manuale Utente Pagina 40

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 172
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 39
3 Making Measurements
30 N9340A User’s Guide
Resolving Small Signals
Hidden by Large Signals
This example uses narrow resolution bandwidths to
resolve two input signals with a frequency
separation of 50 kHz and an amplitude difference
of 60 dB.
1 Connect two sources to the analyzer input
connector as shown in Figure 3 on page 28.
2 Set one source to 300 MHz at –10 dBm. Set the
other source to 300.05 MHz at –70 dBm.
3 Set the analyzer as follows:
Press [PRESET]
Press [FREQ] > {Center Freq} > 300.05 > {MHz}
Press [SPAN] > [500] > kHz.
Press [BW/SWP] > {RBW} > 300 > {Hz}
4 Reduce the resolution bandwidth filter to view
the smaller hidden signal. Place a delta marker
on the smaller signal:
Press [MARKER] > {Peak Search}
Press [MARKER] > {Delta}
Press [MARKER] > {Peak Search} > {Next Right PK}
or {Next Left PK}.
Figure 5 Resolving small signal hidden by a large signal
Vedere la pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 171 172

Commenti su questo manuale

Nessun commento