7 Error Messages
146 N9340A User’s Guide
162 Cannot open fast
sweep in this
case
Indicates current settings do not
allow you to open fast sweep.
For example, currently in FFT,
zero span, or any measurement
in power suite does not allow
fast sweep.
156 Incorrect
alignment file
on flash
Indicates an invalid
alignment file.
153 RF EEPROM
operate fail
Indicates an error when writing
alignment file.
152 Alignment file
oversize
Indicates the alignment file has
error or invalid data.
151 DSP boot fail Indicates the analyzer
cannot process
measurement currently,
wait the analyzer to reboot.
If analyzer froze, try to restart
the analyzer.
150 Mixer overload Indicates the first mixer in
danger. Either increase input
attenuation or decrease the input
signal level.
149 Mixer saturate Indicates you need to either
increase the input attenuation or
decrease the input signal level.
147 Incorrect
alignment data
in EEPROM
Indicates invalid alignment data
occurred.
143 Final IF
overload
Indicates you need to either
increase the input attenuation or
decrease the input signal level.
138 USB device NOT
ready
Indicates the USB device is not
detected.
136 RBW limit to
30 kHz when in
fast sweep
Indicates the analyzer
automatically couple the RBW to
30 kHz when in fast sweep
mode.
Error Code Error Message Description