3 Making Measurements
38 N9340A User’s Guide
Press [MARKER] > {Peak Search}.
Press [MARKER] > {Marker To} > {To Center}.
4 Change the span to 50 MHz and re- center the
signal:
Press [SPAN] > 50 > {MHz}.
Press [MARKER] > {Peak Search}.
5 Set the attenuation to 0 dB:
Press [AMPTD] > {Attenuation} > 0 > {dB}.
Press [MARKER] > {Marker To} > {To Ref}.
6 To determine whether the harmonic distortion
products are generated by the analyzer, first save
the trace data in trace 2 as follows:
Press [TRACE] > {Trace (2)}.
Press [TRACE] > {Clear Write}.
7 Allow trace 2 to update (minimum two sweeps),
then store the data from trace 2 and place a
delta marker on the harmonic of trace 2:
Press [TRACE] > {View}.
Press [MARKER] > {Peak Search}.
Press [Marker] > {Delta}.
Please refer to figure below shows the stored data
in trace 2 and the measured data in trace 1. The
Marker indicator reads the difference in amplitude
between the reference and active markers.
Vedere la pagina 47