4System Setting
66 N9340A User’s Guide
Perform a Time Base Calibration
As soon as the calibration function triggers, it
interrupts the current measurement is interrupted
and a gauge displays on the LCD. The gauge simply
indicates calibration action rather than calibration
course, as the calibration time is unpredictable.
When the calibration is finished, the LCD displays
a calibration, and the interrupted measurement
restarts.
Key Access: [SYS] > {More (2 of 3)} > {Calibration}
NOTE
Time base calibration takes during a short time only when the
inner temperature is stable. When the inner temperature is
increasing, calibration takes a long-time course or will fail. If
the input reference signal is abnormal, the calibration cycle
will take a long and unpredictable time to exit, and the LCD
displays an error message.