Getting Started 2
N9340A User’s Guide 23
Upgrading Firmware
A N9340A provides an easy and fast access for
firmware update.
1 Set up a folder named N9340DATA in the root
directory of your USB memory stick.
2 Store the firmware update application into the
folder named N9340DATA.
3 Insert the USB memory stick into the USB
connector on the top panel of your analyzer.
4 Press [SYS] > {More (2 of 3)} > {Update} >
{Firmware} to activate the firmware updating
procedure. Then the instrument will then
perform the update automatically.
Adding an option
Pressing [SYS] > {More (2 of 3)} > {Option} > {Add}
brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and use the [Enter] key as a
terminator. If the analyzer recognises the option
license code, an message “Option install
successfully” appears in the status line, otherwise a
message “Invalid option licence” appears in the
status line.
CAU-CAUTION
When updating firmware, make sure there will be a
constant power supply for at least 10 minutes. Power cut
off during the updating process can cause damage to the
instrument.