Agilent Technologies N9340A Manuale Utente Pagina 147

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 172
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 146
SCPI Command Reference 6
N9340A User’s Guide 137
Storing as reference
:TGENerator:NORMalize:REF
This command stores the current trace to Trace 4
as a reference. This command is valid after track-
ing generator is enabled.
Normalization
:TGENerator:NORMalize:STATe OFF|ON|0|1
:TGENerator:NORMalize:STATe?
This command toggles the normalization between
on and off.
Normalization Reference Level
:TGENerator:NORMalize:LEVel <para>
:TGENerator:NORMalize:LEVel?
This command sets the normalization level.
Key Access: {Normalize} > {Store Ref}
*RST State: OFF
Key Access: {Normalize} > {Normalize} (On/Off)
*RST State: 0.00 dB
Range: –327.6 to 327.60 dB
Min. Increment 0.01 dB
Available unit: dB
Key Access: {Normalize} > {Norm Ref Lvl}
Vedere la pagina 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 171 172

Commenti su questo manuale

Nessun commento