6 SCPI Command Reference
114 N9340A User’s Guide
Trace Y-Ax is Reference Leve l
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel
<ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
This command sets the amplitude value of the
reference level for the y- axis.
Trace Y-Axis Reference Level Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:
OFFSet<ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:
OFFSet?
This command sets the amplitude level offset for
the Y- Axis.
*RST value: 0.00 dBm
Range: –120.00 to +30 dBm
(in factory default state)
Available unit: dBm, –dBm, mV, mV
Key access: [AMPTD] > {Ref Level}
*RST value: 0.00 dB
Range: –327.60 to +327.60 dB
Available unit: dB
Key access: [AMPTD] > {Ref Offset}