Agilent Technologies 86100D Manuale Utente Pagina 21

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 36
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 20
21
Modules specifications
TDR system
TDR system
(Mainframe with 54754A module) Oscilloscope/TDR performance Normalized characteristics
Rise time 40 ps nominal
< 25 ps normalized
Adjustable from larger of 10 ps or
0.08 x time/div
Maximum: 5 x time/div
TDR step flatness ≤ ±1% after 1 ns from edge
≤ ±5%, –3% < 1 ns from edge
≤ 0.1%
Low level
High level
0.00 V ±2 mV
±200 mV +2 mV
TDR system
(Mainframe with N1055A module)
N1055A - 3xx (35 GHz
bandwidth)
N1055A - 5xx (50 GHz
bandwidth)
Without TDR
calibration
With TDR
calibration
Without TDR
calibration
With TDR
calibration
Step rise/fall time
(Transmitted, typical)
18 ps Adjustable from
18 ps
9 ps Adjustable from
8 ps
86100D Option 202 enhanced impedance and S-parameter software characteristics
Return loss Attenuation
Return loss uncertainty – magnitude
GHz
dB
3
2
1
0
–1
–2
6 dB
6 dB
12 dB
12 dB
20 dB
20 dB
26 dB
26 dB
3 6 9 12 16
–20
–25
–30
–35
–40
–45
–50
–55
–60
16 avgs
64 avgs
256 avgs
0 3 6 9 12 16
Return loss dynamic range – external
GHz
dB
–10
–20
–30
–40
–50
–60
16 avgs
64 avgs
256 avgs
0 4 8 12 16 20 24 28 32
Attenuation uncertainty – magnitude
GHz
dB
2
1
0
–1
–2
–3
6 dB
6 dB
12 dB
12 dB
20 dB
20 dB
30 dB
30 dB
40 dB
40 dB
3 6 9 12 16
Attenuation dynamic range – internal
GHz
dB
–20
–25
–30
–35
–40
–45
–50
–55
–60
16 avgs
64 avgs
256 avgs
0 3 6 9 12 16
Attenuation dynamic range – external
GHz
dB
–10
–20
–30
–40
–50
–60
16 avgs
64 avgs
256 avgs
0 4 8 12 16 20 24 28 32
Vedere la pagina 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 35 36

Commenti su questo manuale

Nessun commento