24
Modules specifications
clock recovery
Clock recovery modules 83496B-100 83496B-101
Electrical through-path digital
amplitude attenuation
5
7.5 dB (No electrical data output signal path)
Wavelength range 830-860 nm and 1260-1360 nm multimode
1260-1360 nm and 1490-1600 nm single mode
Electrical: 150 m Vpp
Front panel recovered
clock output amplitude
1 Vpp max, 220 mVpp min, 300 mVpp
Consecutive identical digits (CID)
150 max
Front panel recovered clock output
divide ratio (user selectable)
6
N=1 to 16 @ data rates 50 Mb/s to 7.1 Gb/s
N=2 to 16 @ data rates 7.1 Gb/s to 14.2 Gb/s
Data input/output connectors 3.5 mm male FC/PC
7
9/125 μm single-mode optical
FC/PC
7
62.5/125 μm multimode optical
3.5 mm male electrical (input only)
Front panel recovered
clock output connector
SMA
1. To convert from OMA to average power with an extinction ratio of 8.2 dB use: PavgdBm = OMA
dBm
–1.68 dB.
2. Verified with PRBS7 pattern, electrical inputs > 150 mVp-p and optical inputs > 3 dB above specification for minimum input level to acquire lock. Output jitter verification
results of the 83496B can be affected by jitter on the input test signal. The 83496B will track jitter frequencies inside the loop bandwidth, and the jitter will appear on the
recovered clock output. Vertical noise (such as laser RIN) on the input signal will be converted to jitter by the limit amplifier stage on the input of the clock recovery. These
effects can be reduced by lowering the Loop bandwidth setting.
3. At rates below 1 Gb/s, loop bandwidth is fixed at 30 KHz when Option 300 is not installed.
4. Without Option 200 loop bandwidth is adjustable from 15 KHz to 6 MHz. Available loop bandwidth settings also depend on the data rate of the input signal. For transition
density from 0.25 to 1, the Loop Bandwidth vs Rate chart shows available loop bandwidth settings. Higher loop bandwidths can be achieved when average data transition
density is maintained at or above 50%.
5. 20*log(Vamp
out
/Vamp
in
) measured with PRBS23 at 14.2 Gb/s.
6. Minimum frequency of divided front panel clock output is 25 MHz.
7. Other types of optical connectors are also available.
Selectable Loop Bandwidth vs Rate
for 0.25 Transition Density 1
10.0E+3
100.0E+3
1.0E+6
10.0E+6
10.0E+6 100.0E+6 1.0E+9 10.0E+9 100.0E+9
Input Data Rate (bits/s)
Loop Bandwidth (Hz)
min
max
DCA plug-in modules (for typical
1
applications)
Electrical / PLL
1 to 12 Gb/s
86112A
Dual channels
BW>20 GHz each
83496B
Electrical clock
recovery
(#100, 200, 300)
Electrical / PLL
1 to 16/32 Gb/s
(high performance)
86108B
Dual channels
BW > 35/50 GHz
with integrated
precision timebase
& clock recovery
Electrical
20, 40 Gb/s
(high-performance)
86118A
Dual remote heads
BW>70 GHz each
86107A or
86100D-PTB
precision timebase
(#40)
Optical
1 to 12 Gb/s
86105C
9 GHz optical channel
20 GHz electrical channel
Optical
20, 40 Gb/s
86105D-281
34 GHz optical channel
(780-1630 nm) 50 GHz
electrical channel
86116C
65 GHz optical channel
80 GHz electrical channel
86107A (#40) or
86100D-PTB
precision timebase
TDR/TDT
Single-ended,
differential
54754A
Dual 18 GHz channels
N1055A
2 or 4 channels
35 or 50 GHz channels
Note - TDR/TDT
modules also perform
as electrical-only
receivers
Trigger options (select one only):
STR – Standard – for basic eye measurements only
ETR – Enhanced – for pattern waveforms, jitter analysis
PTB – Integrated precision timebase for low residual jitter
Remote connection options (select one only):
GPI – GPIB card interface installed
GPN – No GPIB card
86100D hardware options
061/062 – Add Matlab analysis package
200 – Enhanced jitter analysis
201 – Advanced waveform analysis
202 – Enhanced impedance and S-parameters
300 – Advanced amplitude analysis / RIN /Q-scale
500 – Productivity package
SIM – InfiniiSim-DCA de-embedding / embedding
86100DU – 400 PLL and jitter spectrum analysis
86100DU – 401 advanced eye analysis (jitter on PRBS31)
86100D software options (select any):
86100D Infiniium DCA-X mainframe
Typical system configurations
1. Contact your local Agilent sales representative to help configure a system for your specific application.