106
Table 35: Test Port Input
Description Specification Supplemental
Standard Opt 014 Opt UNL Opt 014/UNL
Test Port Noise Floor
a
10 Hz IF Bandwidth
10 MHz to
45 MHz
b
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
--
45 MHz to
500 MHz
c
<-89 dBm <-89 dBm <-89 dBm <-89 dBm --
500 MHz to
2 GHz
<-114 dBm <-114 dBm <-114 dBm <-114 dBm --
2 GHz to
10 GHz
<-117 dBm <-117 dBm <-117 dBm <-117 dBm --
10 GHz to
20 GHz
<-120 dBm <-119 dBm <-120 dBm <-119 dBm --
20 GHz to
40 GHz
<-114 dBm <-113 dBm <-114 dBm <-113 dBm
40 GHz to
50 GHz
<-114 dBm <-112 dBm <-114 dBm <-112 dBm
Option 016
degrades
performance by
2 dB.
1 KHz IF Bandwidth
10 MHz to
45 MHz
b
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
--
45 MHz to
500 MHz
c
<-69 dBm <-69 dBm <-69 dBm <-69 dBm --
500 MHz to
2 GHz
<-94 dBm <-94 dBm <-94 dBm <-94 dBm --
2 GHz to
10 GHz
<-97 dBm <-97 dBm <-97 dBm <-97 dBm --
10 GHz to
20 GHz
<-100 dBm <-99 dBm <-100 dBm <-99 dBm --
20 GHz to
40 GHz
<-94 dBm <-93 dBm <-94 dBm <-93 dBm
40 GHz to
50 GHz
<-94 dBm <-92 dBm <-94 dBm <-92 dBm
Option 016
degrades
performance by
2 dB.