90
Table 30. P11644A Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E836xB/C analyzers, P11644A (WR-62) calibration kit, 85132F flexible test port cable set with the P281B and P281C launch
sets, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
12.4 to
18 GHz
Directivity 50
Source Match 50
Load Match 50
Reflection Tracking
±0.000
+0.02/°C
Transmission Tracking
±0.016
+0.02/°C
Commenti su questo manuale