81
E8363B/C AND E8364B/C Corrected System Performance with WR-42 Connectors
Table 27. K11644A Calibration Kit
Standard Configuration and Standard Power Range
(E8363B/C AND E8364B/C)
Applies to the, E8363B/C AND E8364B/C analyzers, K11644A (WR-42) calibration kit, 85134F flexible test port cable set with the K281C
launch set,
, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
18 to
20 GHz
20 to
26.5 GHz
Directivity 50 50
Source Match 50 50
Load Match 50 50
Reflection Tracking
±0.000
+0.02/°C
±0.000
+0.02/°C
Transmission Tracking
±0.014
+0.02/°C
±0.018
+0.02/°C