Agilent Technologies E8364C Manuale Utente Pagina 13

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 145
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 12
9
E8363B/C AND E8364B/C Corrected System Performance with 2.4mm Connectors
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
Table 3. 85056A Calibration Kit
Standard Configuration and Standard Power Range
(E8363B/C AND E8364B/C)
Applies to the E8363B/C AND E8364B/C analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
40 GHz
40 to
50 GHz
Directivity 42 42 38 36
Source Match 41 38 33 31
Load Match 42 42 37 35
Reflection Tracking
±0.001
+0.02/°C
±0.008
+0.02/°C
±0.020
+0.02/°C
±0.027
+0.03/°C
Transmission Tracking
±0.010
+0.02/°C
±0.049
+0.02/°C
±0.105
+0.02/°C
±0.170
+0.03/°C