84
Table 28. K11644A Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E8363B/C AND E8364B/C analyzers, K11644A (WR-42) calibration kit, 85134F flexible test port cable set with the K281C
launch set,, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
18 to
20 GHz
20 to
26.5 GHz
Directivity 50 50
Source Match 50 50
Load Match 50 50
Reflection Tracking
±0.000
+0.02/°C
±0.000
+0.02/°C
Transmission Tracking
±0.016
+0.02/°C
±0.019
+0.02/°C