
125
Table 45. Frequency Converter Application (option 083) Cycle Time for Fixed-IF Measurements (ms)
1
Number of Points
101 201 401
Stimulus start = 1 GHz, stop = 11 GHz, IFBW = 35 kHz
Response = 70 MHz, trace = SC21, cal = SMC_2P
Hardware Trigger 8.5 17 34
Software Trigger 31 62 124
Table 46. Data Transfer Time (ms)
a
Number of Points
201 401 1601 16,001
SCPI over GPIB
(program executed on external PC)
32-bit floating point 7 12 43 435
64-bit floating point 12 22 84 856
ASCII 64 124 489 5054
SCPI
(program executed in the analyzer)
32-bit floating point 1 2 3 30
64-bit floating point 2 2 4 40
ASCII 29 56 222 2220
COM (program executed in the analyzer)
32-bit floating point 1 1 1 6
Variant type 1 2 6 68
DCOM over LAN
(program executed on external PC)
32-bit floating point 1 1 2 121
Variant type 3 6 19 939
a
Typical performance
Note: Specifications for Recall & Sweep Speed are not provided for the E836xB/C analyzers.
Commenti su questo manuale