21
E8363B/C AND E8364B/C Corrected System Performance with 2.92mm Connectors
Table 7. 85056K Calibration Kit
Standard Configuration and Standard Power Range
(E8363B/C AND E8364B/C)
Applies to the, E8363B/C AND E8364B/C analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
0.045 to
2 GHz
2 to
20 GHz
20 to
40 GHz
Directivity 42 42 40
Source Match 40 40 35
Load Match 42 41 38
Reflection Tracking
±0.018
+0.02/°C
±0.018
+0.02/°C
±0.067
+0.03/°C
Transmission Tracking
±0.011
+0.02/°C
±0.042
+0.02/°C
±0.089
+0.03/°C