Agilent Technologies E8364C Manuale Utente Pagina 43

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 145
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 42
39
Table 13. 85052D Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85052D (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Data
and traces above 20 GHz are not applicable to the E8362C. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
26.5 GHz
Directivity 42 36 30
Source Match 37 28 25
Load Match 42 36 30
Reflection Tracking
±0.003
+0.02/°C
±0.008
+0.02/°C
±0.011
+0.03/°C
Transmission Tracking
±0.014
+0.02/°C
±0.131
+0.02/°C
±0.250
+0.03/°C
Vedere la pagina 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 144 145

Commenti su questo manuale

Nessun commento