Agilent Technologies E8364C Manuale Utente Pagina 115

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 145
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 114
111
Table 35 (Continued). Test Port Input
Trace Noise Magnitude - Option 080 enabled
b,d
1 kHz IF bandwidth. Ratio measurement, nominal power at test port. Typical
10 MHz to
45 MHz
-- <0.060 dB rms
45 MHz to
500 MHz
e
-- <0.010 dB rms
500 MHz to
20 GHz
-- <0.006 dB rms
20 GHz to
40 GHz
-- <0.007 dB rms
40 GHz to
50 GHz
-- <0.008 dB rms
Trace Noise Phase
1 kHz IF bandwidth. Ratio measurement, nominal power at test port.
10 MHz to
45 MHz
b
<0.350° rms (typical) --
45 MHz to
500 MHz
<0.100° rms --
500 MHz to
20 GHz
<0.060° rms --
20 GHz to
40 GHz
<0.100° rms --
40 GHz to
50 GHz
<0.100° rms --
Trace Noise Phase - Option 080 enabled
b,d
1 kHz IF bandwidth. Ratio measurement, nominal power at test port.
Typical
10 MHz to
45 MHz
-- <0.350° rms
45 MHz to
500 MHz
e
-- <0.100° rms
500 MHz to
20 GHz
-- <0.060° rms
20 GHz to
40 GHz
-- <0.100° rms
40 GHz to
50 GHz
-- <0.100° rms
Reference Level Magnitude
Range +/-200 dB --
Resolution 0.001 dB --
Reference Level Phase
Range +/-500° --
Resolution 0.01° --
Vedere la pagina 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 144 145

Commenti su questo manuale

Nessun commento