Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
10
4 Set up the [Channel] screen as follows:
On the 81134A set up both channels and switch on only the channel
that is being tested.
5 Set the Digitizing Oscilloscope Agilent 54121T:
Press [AUTOSCALE]
Select the Display menu and set the Number of Averages to 32
Select the Delta V menu and switch the voltage markers on
Set the preset levels to 50% - 50% and press [AUTO LEVEL SET]
Select the Delta T menu and switch the time markers on
Set START ON EDGE = POS 1 and STOP ON EDGE = NEG 1
6 Change the oscilloscope timebase to 100 ps/div
7 Change the Agilent 81133A/81134A Width to 100 ps
8 Center the pulse in the oscilloscope display
9 Press the [PRECISE EDGE FIND] key for each new width setting