Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
24
On the 81134A set up both channels and switch on only the channel that
is being tested.
4 Write down the measured voltage for the "zero Volt" level.
5 On the "Channel" screen switch to "Inverted" Levels and measure the
high level.
6 Calculacte the Amplitude from the measured levels and repeat the test
for the following high levels, always measuring additionially the low
level as well
High Level Acceptable Range TR Entry
50 mV 39 mV to 61 mV 5-11
1.00 V 0.97 V to 1.03 V 5-12
2.00 V 1.95 V to 2.05 V 5-13
For the 81134A repeat the test for the second channel.