Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
7
Test 1: Frequency
Range: 15 MHz to 3.35 GHz
Resolution: 1 Hz
Accuracy: 50 ppm
Counter
Cable 50 Ω, SMA
1 Connect the 81133A/81134A Trigger Output to the Counter Channel 1
Input
2 Reset the 81133A/81134A to default settings
3 Select the [Aux] screen on the 81133A/81134A and set up the unit as
follows: