Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
12
Test 3: Delay
Range: Variable Delay - 5 ns to + 230 ns
Resolution: 1 ps
Accuracy: ± 20 ps after autocal within a temperature range of ± 10°
Digitizing Osccilloscope with Accessories
Power Splitter
Cables 50 Ω, SMA
1 Connect the 81133A/81134A to the oscilloscope as shown:
SMA-SMA Adaptor
Trig Out
Channel 2 Channel 1
Output
Input 1 2 3 4 TRIG
Power Splitter
20 dB Attenuator
Commenti su questo manuale