Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
30
6 Negative Edge Overshoot
On the Oscilloscope
Select the Channels menu and set the channel 3 Offset to 0 V and
VOLTS/DIV to 500 mV/div
Select the Timebase menu and set the Variable Levels to -10% and 10%
Press Auto Level Set
Select the Timebase menu and set TIME/DIV to 500 ps7div
Adjust the delay to center the negative edge
Select the Channels menu and set the channel 3 offset to - 0.95 V
Set VOLTS/DIV to 200 mV/div
7 Check that the Overshoot of the negative edge is within the ± 10 % of
the amplitude limits and record the result in TR Entry 7-2
The maximum specified overshoot is actually ± 10 % ± 10 mV which is
± 220 mV for an amplitude of 2 V. If the Overshoot is noth within the
± 10% of amplitude limits, make a precise measurement against the
specification of ± 220 mV.
For the 81134A repeat the test for channel 2.
Important
Vedere la pagina 29