Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
14
3 Set the Digitizing Oscilloscope Agilent 54121T:
• Press [AUTOSCALE]
• Set the Timebase to 5 ns/div
• Adjust the Delay to 25 ns
• Select the Display menu and set the screen function to single; set the
number of averages to 16
• Select the Delta V menu and switch the Volatge Markers on
• Assign Marker 1 to Channel 3 and Marker 2 to Channel 4
• Set the Preset Levels to 50% - 50% and press Auto Level Set
• Select the Delta t menu and switch the time markers on
• Set START ON EDGE = POS 1 and STOP ON EDGE = POS 1
• Press the [Precise Edge Find] key
4 Record the measured value on the Test Record as the Fixed Delay ∆t
0
for the Channel under test
5 Set the 81133A/81134A Delay to 5 ns, then press [CLEAR DISPLAY]
and the [Precise Edge Find] key on the scope
6 Check the Delay at the following settings
Oscilloscope
Timebase
Delay Acceptable Range TR Entry
200 ps/div 1 ns 980 ps to 1.02 ns 3-1
500 ps/div 1.5 ns 1.48 ns to 1.52 ns 3-2
500 ps/div 2 ns 1.98 ns to 2.02 ns 3-3
Larger delays are generated as counted delay. The performance is
guaranteed by the design and does not need to be verified.
For the 81134A repeat the test for the second channel.