Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
9
Test 2: Width
Range: 100 ps to (period - 100 ps)
Resolution: 1 ps
Accuracy: ± 40 ps after autocal within a temperature range of ± 10°
Digitizing Oscilloscope with Accessories
Cables 50 , SMA
1 Connect the 81133A/81134A Trigger Output to the Oscilloscope
Trigger Input, using a 20 dB attenuator
2 Connect the 81133A/81134A Channel 1 Output to the Oscilloscope
Channel 4 Input, using a 20 dB attenuator
3 Set up the Pulse Generator as follows:
Equiopment Needed
Procedure