Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
19
Test 5a: High and Low Levels
Amplitude: 50 mV
pp
to 2 V
pp
Level Window: -2.00 V to +3.00 V
Level Resolution: 10 mV
Level Accuracy: 2 % of setting ± 20 mV
Digitzing Voltmeter
50 Ω Feedthrough Termination, 0.1 %, 10 W
BNC to Dual Banana plug
SMA Cables 50 Ω, SMA/BNC Adaptor
1 Connect the 50 Ω Feedthrough Termination to the DVM input, using
the BNC to Dual Banana plug
2 Connect the 81133A/81134A channel output to the 50 Ω Feedthrough
Termination, using the SMA/BNC adaptor
Commenti su questo manuale