Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
26
On the 81134A set up both channels and switch on only the channel that
is being tested.
4 Rise Time 10% - 90%
On the Oscilloscope:
Press [AUTOSCALE]
Center one pulse on the screen
Select the Display menu and set the Number of Averages to 32, set the
bandwidth to 20 GHz
Select the Channel menu and set the Attenuation Factor of channel 2
to 10
Select the Delta V menu and switch on the V markers
Set the Preset Levels to 10 - 90 % and press Auto Level Set
Select the Timebase menu and set TIME/DIV = 20 ps/div
Adjust the delay until a positive edge is centered
Select the delta t menu and switch on the t markers
Set START ON EDGE = POS 1 and STOP ON EDGE = POS 1
Press the Precise Edge Find key
5 Check that the Rise Time 10% - 90% is < 90 ps and record the measured
valued in TR entry 6-1