Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
6
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following figure provides a schematic and a parts list except for the
case. The case must provide shielding and maintain grounding integrity.
50 Ohm, 0.1%, 10W Feedthrough Termination
The following parts are required:
• R1 = 53.6 Ω, 1%, 10 W
• R2 = 200 Ω, 10%, 0.5 W, Variable Trimmer
• R3 = 681 Ω, 1%, 0.5 W
• Connectors: BNC male, BNC female