Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
18
12 Set up the Digitizing Oscilloscope Agilent 54121T:
• Press [AUTOSCALE]
• Select the Display menu and set the Number of Averages to 64
• Select the Channel menu and set the Attenuation factor of channel 2 to
2
• Set the VOLTS/DIV of channel 2 to 10 mV/div, the OFFSET to 500 mV
• Select the Timebase menu and set the TIME/DIV to 10 ps/div
13 Center a positive edge of the signal
14 Press [MORE] and [HISTOGRAM]
• Select the Window submenu and set:
• Source is channel 2
• Choose the Time Histogram
• Press [WINDOW MARKER 1] and set it to 490 mV
• Press [WINDOW MARKER 2] and set it to 500 mV
15 Select the Acquire submenu, set the Number of Samples to 1000 and
press [START ACQUIRING]
16 After the data for the time histogram has been acquired (#Samples =
100 %) select the Result submenu
17 Press [MEAN] and [SIGMA]. Record the values of sigma as the RMS
jitter.
18 Check the data jitter at the following frequencies:
Frequency Acceptable Range TR Entry
100 MHZ < 4 ps 4-5
1 GHz < 4 ps 4-6
2 GHz < 4 ps 4-7
19 Set the Channel Frequency Divider of the 81133A/81134A to 2 and
check the data jitter at 3 GHz. Record the result in TR Entry 4-8.
For the 81134A repeat the tests for channel 2.