Vedere la pagina 10
Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
11
10 Check the Agilent 81133A/81134A pulse width at the following
settings:
Oscilloscope
Timebase
Width Acceptable Range TR Entry
100 ps/div 100 ps 60 ps to 140 ps 2-1
100 ps/div 500 ps 460 ps to 540 ps 2-2
500 ps/div 1 ns 960 ps to 1.040 ns 2-3
500 ps/div 2 ns 1.96 ns to 2.04 ns 2-4
Larger pulse widths are generated with counted delays. The performance
is guaranteed by the design and does not need to be verified.
For the 81134A repeat the test for the second channel.
Vedere la pagina 10