Power Supply
Voltages Check
The power supply must be loaded by either the System Assembly Board or with an
added resistor to check the voltages.
Warning This procedure is to be performed only by service-trained personnel
aware of the hazards involved (such as fire and electrical shock).
Power Supply Loaded by System Assembly
1. Remove instrument top cover.
2. Using the figure below, check for the voltages indicated at the testpoints.
Power Supply Isolated
Isolate and check the supply with the following steps. Use the figure above for
reference.
1. Remove instrument power cable.
2. Disconnect supply output cable at supply (see figure above).
3. Load + 5.20 V supply with a 2 ohm 25 watt resistor. Use jumper wires to
connect one end of the resistor to any of pins 1-4, and the other end to any of
pins 5-8.
Figure 6-11. Power Supply Test Points
Service HP 1650B/1651B
6-14 Service Manual