2. Adjust pulse generator for the output in figure 3-11. Pulse generator settings
are different for HP 1650B and HP 1651B.
Setting for HP 8161A:
Parameter Output A Output B
Input Mode Norm ---
Period (PER)
HP 1650B 28.5 ns ---
HP 1651B 40 ns ---
Width (WID)
HP 1650B 18.5 ns 18.5ns
HP 1651B 30 ns 30 ns
Leading Edge (LEE) 1 ns 1 ns
Trailing Edge (TRE) 1 ns 1 ns
High Level (HIL) 3.2 V 3.2 V
Low Level (LOL) 0 V 0 V
Delay (DEL) 0 ns 0 ns
Output Mode ENABLE ENABLE
3. Assign the pod under test to Analyzer 1 in System Configuration as in
previous test figure 3-4.
4. In the State Format Specification assign Clock Period < 60 ns, and rising
edge of J clock. Also, assign lower 8 channels of the pod under test to a label
as in figure 3-12.
Figure 3-11. Waveform for Data Test 2
Figure 3-12. State Format Specification for Data Test 2
Performance Tests HP 1650B/1651B
3-10 Service Manual