Clock,
Qualifier, and
Data Inputs
Test 1
Description: This test verifies maximum clock rate with counting mode for the HP
1650B/51B. This test also verifies the setup and hold times for the falling edge of
the HP 1650B L clock specification, and the falling edge of the HP 1651B J and K
clock specification.
Specification: Clock repetition rate: with time or state counting mode on,
minimum time between states is 60 ns.
Setup time: Data must be present prior to clock transition;
10 ns.
HP 1650B hold time: Data must be present after falling edge of L clock transition;
0 ns.
HP 1651B hold time: Data must be present after falling edge of J and K clock
transition; 0 ns.
Equipment:
Pulse Generator .............................. HP 8161A/020
Oscilloscope .................................... HP 54502A
50 Ohm Feedthru (2) ............................. HP 10100C
Test Connector see figure 3-1......................... (2)
Procedure:
1. Connect the HP 1650B\ 51B and test equipment as in figure 3-2.
Note In this setup, only eight channels are tested at a time to minimize loading. Ground
lead must be connected to ensure accurate test results.
Figure 3-2. Setup for Data Test 1
Performance Tests HP 1650B/1651B
3-4 Service Manual