Agilent Technologies 54502A Manuale di Servizio Pagina 18

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 111
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 17
Tagging
State Tagging: Counts the number of qualified states between each stored state.
Measurement can be shown relative to the previous state or relative to trigger.
Maximum count is 4.4 X (10 to the 12th power).
Time Tagging: Measures the time between stored states, relative to either the
previous state or the trigger. Maximum time between states is 48 hours.
With tagging on, the acquisition memory is halved; minimum time between states is
60 ns.
Symbols
Pattern Symbols: User can define a mnemonic for the specific bit pattern of a
label. When data display is SYMBOL, mnemonic is displayed where the bit
pattern occurs. Bit pattern can include 0s, 1s, and don’t cares.
Range Symbols: User can define a mnemonic covering a range of values. Bit
pattern for lower and upper limits must be defined as a pattern of 0s and 1s. When
data display is SYMBOL, values within the specified range are displayed as
mnemonic + offset from base of range.
Number of Pattern and Range Symbols: 200 total used for both analyzers.
Symbols can be down-loaded over RS-232-C or HP-IB.
Timing Analysis Transitional Timing Mode
Sample is stored in acquisition memory only when the data changes. A time tag
stored with each sample allows reconstruction of waveform display. Time covered
by a full memory acquisition varies with the number of pattern changes in the data.
Sample Period: 10 ns.
Maximum Time Covered By Data: 5000 seconds.
Minimum Time Covered by Data: 10.24 us.
Glitch Capture Mode
Data sample and glitch information stored every sample period.
Sample Period: 20 ns to 50 ms in a 1-2-5 sequence dependent on sec/div and delay
settings.
Memory Depth: 512 samples/channel.
Time Covered by Data: Sample period X 512.
Waveform Display
Sec/div: 10 ns to 100 s; 0.01% resolution.
Delay: -2500 s to 2500 s; presence of data dependent on the number of transitions
in data between trigger and trigger plus delay (transitional timing).
Accumulate: Waveform display is not erased between successive acquisitions.
HP 1650B/1651B General Information
Service Manual 1-7
Vedere la pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 110 111

Commenti su questo manuale

Nessun commento