Keyboard Signals
Check
Isolate a faulty elastomeric keypad or keyboard when random key(s) not operating
by performing the following steps.
1. Remove instrument power cable.
2. Without disconnecting the keyboard cable, follow keyboard removal
procedure to loosen keyboard. Leave keyboard in place in front of
instrument.
3. Apply power.
4. Run Keyboard Self Test and press all keys.
5. Allow keyboard assembly to fall forward from front panel. Separate the
elastomeric keypad and keyboard panel from the PC board.
6. Short PC board trace (with a paper clip or screwdriver) of non-operating key
and look for appropriate response on display.
7. If display responds as if key were pressed, replace elastomeric keypad.
8. If display does not respond as if key were pressed, replace keyboard.
The RPG connector has a TTL pulse on pins 1 and 3, when the knob is being
turned. Pin 5 of the connector is + 5 V.
The ROW (scan) signal is approximately 60 Hz, a low duty-cycle pulse. It is
continually present on pins 11 through 18 of the keyboard cable. Because of the
resistance of the keypad contacts, the signal does not appear the same on the
COLUMN (data) pins when keys are pressed. Refer to the following table for
signals going to and from the keyboard.
PIN SIGNAL PIN SIGNAL
1 RPGA 2 RPGB
3 COLUMN 1 (Data) 4 COLUMN 2 (Data)
5 COLUMN 3 (Data) 6 COLUMN 4 (Data)
7 COLUMN 5 (Data) 8 COLUMN 6 (Data)
9 COLUMN 7 (Data) 10 COLUMN 8 (Data)
11 ROW 8 (Scan) 12 ROW 7 (Scan)
13 ROW 6 (Scan) 14 ROW 5 (Scan)
15 ROW 4 (Scan) 16 ROW 3 (Scan)
17 ROW 2 (Scan) 18 ROW 1 (Scan)
19 +5 V (Digital) 20 GROUND (Digital)
Table 6-3. Keyboard Connector Voltages and Signals
Service HP 1650B/1651B
6-16 Service Manual