
Recommended
Test Equipment
Recommended test equipment for calibration and adjustment is listed in table 1-3.
Any equipment that satisfies the critical specifications given in the table may be
substituted for the recommended models.
Warning Read the safety summary at the front of this manual before any
adjustment, replacement, maintenance, or repair is performed.
Instrument
Warmup
Adjust and calibrate the instrument at it’s environmental ambient temperature and
after a 15 minute warm-up.
Calibration The calibration procedures in this section should be followed in their entirety and
in the same sequence as in this section.
Adjustments HP 1650B/1651B
4-2 Service Manual
Commenti su questo manuale