10. Align new fan on back panel with fan cable adjacent to line filter assembly
and label toward rear (fan rotor is facing forward).
11. Align fan guard on back of panel.
12. Attach fan to rear panel with four machine screws.
13. Re-assemble instrument in reverse order of above.
14. Install top cover. Secure top cover to chassis using ten screws.
Note To ensure top cover is properly installed, secure side screws of top cover before
securing top screws.
Removal and
Replacement of
Keyboard
Assembly
1. Disconnect pod cables from rear of instrument.
2. Disconnect power cable.
3. Remove ten screws securing top cover. Remove top cover.
4. Disconnect disk drive cable from disk drive.
5. Remove two screws securing disk drive to power supply. Remove disk drive
by sliding assembly out front panel of instrument.
6. Remove two retaining pins securing power supply to chassis. Slide power
supply out far enough to gain unobstructed access to power supply cables.
7. Disconnect power supply line input and DC output cables. Remove power
supply.
8. Disconnect keyboard cable from system board.
9. Remove four screws securing keyboard assembly to chassis (located on
inside of instrument).
10. Remove keyboard assembly.
11. Disconnect RPG cable from keyboard circuit board assembly. Keyboard
assembly can now be disassembled.
12. Reinstall keyboard assembly by reversing this procedure.
13. Install top cover. Secure top cover to chassis using ten screws.
Note To ensure top cover is properly installed, secure side screws of top cover before
securing top screws.
HP 1650B/1651B Service
Service Manual 6-21
Commenti su questo manuale