Operating
Characteristics
The following operating characteristics are are typical operating characteristics for
the HP 1650B/51B Logic Analyzer.
Probes Input RC: 100 K 2% shunted by approximately 8 pF at the probe tip.
TTL Threshold Preset: + 1.6 volts.
ECL Threshold Preset: -1.3 volts.
Threshold Range: -9.9 to + 9.9 volts in 0.1V increments.
Threshold Setting: Threshold levels may be defined for pods 1 and 2 individually
(HP 1651B). Threshold levels may be defined for pods 1, 2, and 3 on an individual
basis and one threshold may be defined for pods 4 and 5 (HP 1650B).
Minimum Input Overdrive: 250 mV or 30% of the input amplitude, whichever is
greater.
Maximum Voltage:
40 volts peak.
Maximum Power Available Through Cables: (600 mA) 2/3 amp @ 5V per cable; 2
amp @ 5V per H P 1650B/51B.
Measurement
Configurations
Analyzer Configurations:
Analyzer 1 Analyzer 2
Timing Off
Off Timing
State Off
Off State
Timing State
State Timing
State State
Off Off
Channel Assignment: Each group of 16 channels (a pod) can be assigned to
Analyzer 1, Analyzer 2, or remain unassigned. The HP 1650B contains 5 pods; the
HP 1651B contains 2 pods.
HP 1650B/1651B General Information