Agilent Technologies 54502A Manuale di Servizio Pagina 43

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 111
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 42
6. Press RUN. The State Listing is displayed and shows Fs for the channels
under test, when test is passed, as in figure 3-9.
Note To ensure consistent pattern of Fs in listing, use front-panel ROLL field and knob
to scroll through State Listing.
7. If testing the HP 1651B, connect the K clock of Pod 2 to the test connector
and repeat steps 4 and 6 for the falling edge of K clock.
8. Remove the probe tip assembly from the logic analyzer probe cable and
attach to the next logic analyzer probe cable to be tested. Take care not to
dislodge grabbers from the test connector. If testing the HP 1651B then
re-assign the falling edge of J clock.
9. Repeat steps 3, 4, 6 and 7 until all pods have been tested.
10. Disconnect lower eight bits (bits 0 through 7) from test connector. Attach
upper eight bits (bits 8 through 15) to the test connector.
11. Repeat steps 3, 4, 6, 7 and 8 until the upper bits of all pods have been tested.
Figure 3-9. State Listing for Data Test 1
Performance Tests HP 1650B/1651B
3-8 Service Manual
Vedere la pagina 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 110 111

Commenti su questo manuale

Nessun commento