2. Adjust pulse generator for the output in figure 3-26.
Setting for HP 8161A:
Parameter Output A Output B
Input Mode Norm ---
Period (PER) 120 ns ---
Width (WID) 60 ns 10 ns
Leading Edge (LEE) 1 ns 1 ns
Trailing Edge (TRE) 1 ns 1 ns
High Level (HIL) 3.2 V 3.2 V
Low Level (LOL) 0 V 0 V
Delay (DEL) 40 ns ---
Double Pulse (DBL) --- 60 ns
Output Mode ENABLE ENABLE
3. Assign the pod under test to Analyzer 1 in System Configuration as in
previous figure 3-4.
4. Set up State Format Specification as in figure 3-27, assigning the falling J
clock as Master Clock and the rising J clock as Slave Clock. Refer to steps a
through d after figure if unfamiliar with the menus.
a. Move cursor to Clock field and SELECT, then assign Demultiplex.
Figure 3-26. Waveform for Data Test 6
Figure 3-27. State Format Specification for Data Test 6
Performance Tests HP 1650B/1651B
3-22 Service Manual