4. Reconnect instrument power cable and check for voltages at the supply
output using values in the following table.
PIN SIGNAL PIN SIGNAL
1 +5.20 V 11 -5.2 V
2 +5.20 V 12 GROUND
3 +5.20 V 13 +12 V
4 +5.20 V 14 GROUND
5 GROUND (Display) 15 -12 V
6 GROUND (Digital) 16 GROUND
7 GROUND (Digital) 17 +12 V (Display)
8 GROUND 18 -5.2 V
9 +3.5 V 19 +15.5 V (Fan)
10 GROUND 20 GROUND (Fan)
Note The ground planes (digital, fan, and display) are at the same potential on the
power supply, but when measuring on the main assembly the supplies must be
measured with reference to the respective ground.
CRT Monitor
Signals Check
1. Remove instrument top cover.
2. Check the CRT Monitor input cable for the signals and supplies listed in the
table below. The cable is the wide ribbon cable connecting the monitor to
the System Assembly Board.
3. Dynamic video signals FB (Full-bright) and HB (Half-bright) are TTL
inputs. Check for activity on these lines. The table includes a truth table for
these signals.
PIN SIGNAL PIN SIGNAL FB HB VIDEO
1 +5 V (Digital) 2 +12 V (Display) 0 0 OFF
3 GROUND (Display) 4 GROUND (Display) 0 1 HALF
5 +12 V (Display) 6 GROUND (Display) 1 0 FULL
7 +12 V (Display) 8 GROUND (Display) 1 1 FULL
9 +12 V (Display) 10 HSYNC
11 VSYNC 12 +12 V (Display)
13 GROUND (Digital) 14 GROUND (Digital)
15 GROUND (Display) 16 FB (Full-bright)
17 GROUND (Display) 18 HB (Half-bright)
19 GROUND (Display) 20 +5 V (Digital)
Table 6-1. Power Supply/Main Assembly Voltages