8712ET/ES and 8714ET/ES Adjustments
RF Network Analyzers Set Serial Number, Adjustment #100
Service Guide 2-7
Set Serial Number, Adjustment #100
Adjustment #100 cannot be used to set an established serial number. The procedure
outlined below can only be used to enter the serial number if the analyzer displays a
default serial number value of
?????????? (ten question marks).
If it is necessary to modify a serial number that is other than the default value, contact the
nearest Agilent Technologie sales or service office (see Chapter 9).
NOTE Where XXXXXXXXXX appears throughout this procedure, replace those
characters with the serial number of your analyzer, but maintain the leading
and following apostrophes (both are ASCII character 39). For example, if the
serial number of your analyzer is US34405555, the HP BASIC line of code
would be:
OUTPUT 800;"DIAG:SNUM 'US34405555'"
IMPORTANT: Make sure there is a space between SNUM and the first
apostrophe.
Required Equipment
A PC-compatible keyboard is required.
If a keyboard is not available, this adjustment can be done from the front panel. (Refer to
the note in the following procedure.)
Estimated Adjustment Time
This adjustment takes approximately two minutes to perform.
Procedure
1. Write down the 10-character serial number, exactly as shown on the analyzer rear-panel label.
2. With an external keyboard connected, press to view the IBASIC command line. Then
type the following:
OUTPUT 800;"DIAG:SNUM 'XXXXXXXXXX'"
then press .
NOTE If you are using the front panel to set the serial number, enter the following
IBASIC code, then “run” the program.
OUTPUT 800;"DIAG:SNUM 'XXXXXXXXXX'"
END
Refer to Chapter 5 of the HP Instrument Basic User’s Handbook Supplement
for information on programming from the front panel of the instrument.
3.
Press to verify that you correctly
installed the serial number. The serial number is displayed in the dialogue box.
ESC
ENTER
SYSTEM OPTIONS