8712ET/ES and 8714ET/ES Assembly Replacement
RF Network Analyzers A6 Power Supply Assembly Replacement
Service Guide 3-19
A6 Power Supply Assembly Replacement
The power supply is a single assembly. The only lower level part available for replacement
is the cooling fan.
CAUTION Perform the followingonly at an approved ESD workstation. SeeFigure 4-1
on page4-3.
Removing the Power Supply Assembly
1.Disconnect the power cord.
2.Remove the front panel assembly (see“Removing the Front Panel Assembly” on page
3-4). Remove the handle nut plate on the left side only.
3.Remove the display enclosure (see the following section“Removing the Display
Assembly” on page 3-20
4.Remove the four screws along the rear edge of the display enclosure.
5.Remove the screw from the display access panel.
6.Open the display access panel and disconnect the two-wire cable from the power supply.
7.Remove the power supply from the housing.
Installing the Power Supply Assembly
To install the power supply assembly, reverse the steps in the“Removing the Power Supply
Assembly” procedure.