Performance Checks 8712ET/ES and 8714ET/ES
Test Port Input RF Network Analyzers
1-64 Service Guide
Table 1-9 Test Port Input (continued)
1. Noise floor is defined as the RMS value of the trace (in linear
format) for a transmission measurement in CW Frequency
mode, with the RF connectors terminated in loads, output
power set to 0 dBm, and no averaging. The noise floor specifi-
cations and typicals for narrowband detection measurements
assume that an isolation calibration has been performed
using an average factor of 16. For the 8712ES/8714ES, exter-
nal broadband detectors will provide a lower noise floor than
the internal broadband detectors.
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dBm)
Typical
(in dBm)
System Bandwidths:
Fine
(15 Hz)
Fine
(15 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
Noise Floor
1
8712ET/8714ET:
50 :
Narrowband 105 111 86 48
Broadband (Internal) 50 55 48 −32
75 :
Narrowband 104 109 84 46
Broadband (Internal) 47 52 45 30
8712ES/8714ES:
50 Ω:
Narrowband 96 105 80 47
Broadband (Internal) 38 43 36 −23
75 :
Narrowband 95 104 80 47
Broadband (Internal) 35 40 33 −20