Performance Checks 8712ET/ES and 8714ET/ES
Performing the Operator's Check RF Network Analyzers
1-4 Service Guide
Make a Transmission (or S21 and S12) Measurement(s)
1. Connect the equipment as shown in Figure 1-1. Use a known good cable such as the one
that was supplied with your analyzer.
NOTE The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 1-3.
Figure 1-1 Equipment Setup for Performing the Operator’s Check
2. Press .
3. Press .
4. Press .
5. Verify that the data trace falls within ±0.5 dB of 0 dB. See Figure 1-2 for a typical 8714
result. The 8712 should look similar, but will end at 1300 MHz.
PRESET SCALE .1
Enter
POWER 0
dBm
CAL
Default Response