Assembly Replacement 8712ET/ES and 8714ET/ES
A2 CPU Board Assembly Replacement RF Network Analyzers
3-8 Service Guide
Figure 3-3. Location of the BootROM Chip, U335 and Jumper J301
7. Transfer any modified (Single Inline Memory Module) SIMM:
If the original board has a larger SIMM installed, transfer the larger SIMM to the new
CPU assembly.
NOTE As shipped from the factory, all analyzers have 32 MB of SIMM memory. To
increase the amount of memory, remove the existing SIMM and replace it
with a higher capacity SIMM.
8. Check the jumper at J301 (see Figure 3-3) and verify that it is set for your analyzer’s
model type. All ET and ES models should have the jumper on the “E” designation. The
“C” designation is for the previous HP/Agilent 871xC models.
9. Transfer the A8 3.5" internal disk drive assembly:
Remove the disk drive from the old board and install it on the replacement assembly.
10.Transfer all cable assemblies:
Remove all cable assemblies from the old board and properly attach them to the
replacement board.
The replacement CPU board is now ready to be installed.