8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
Service Guide 1-13
File Structure
The program disk contains several different files in addition to the main program file. The
following list describes the files found on the performance test software disk.
CAL871XE This is the main program file.
Cfg_xxxx This is the configuration file. The variable
xxxx represents any
user-defined name or
Dflt (default). This file contains all the configuration
set-ups for the program such as:
file directories and paths
test equipment serial numbers
GPIB addresses
types of test equipment
other optional choices
If more than one configuration file exists on the disk, the program will
prompt you to select a specific configuration file to load. If none exists, the
program will create a default file.
P1D_xxxx The cal factor file for the HP/Agilent 8481D sensor. The variable
xxxx
represents the last four digits of the HP/Agilent 8481D or HP/Agilent
8484A sensor's serial number.
P2A_xxxx The cal factor file for the HP/Agilent 8482A sensor. The variable xxxx
represents the last four digits of the 8482A sensor's serial number.
P3A_xxxx The cal factor file for the HP/Agilent 8483A sensor. The variable
xxxx
represents the last four digits of the 8483A sensor's serial number.
Stp_xxxx The actual attenuation values for the external 110 dB step attenuator. The
variable xxxx represents the last four digits in the attenuator's serial
number. Attenuator values are measured at 30 MHz on an accurate
system (such as an HP/Agilent 8902A Option 050).
Pd5_xxxx The actual value of the 20 dB 50 attenuator as measured by an accurate
system. Values are attenuation versus frequency. The variable
xxxx
represents the last four digits in the attenuator's serial number.
Pd7_xxxx The actual value of the 20 dB 75 attenuator as measured by an accurate
system. Values are attenuation versus frequency. The variable xxxx
represents the last four digits in the attenuator's serial number.
dxxExxxx The data file that holds the analyzer's test results. The variable
xxxx
represents the last four digits in the analyzer's serial number. The file
“d2TExxxx” holds data for an 8712ET, “d4SExxxx” for an 8714ES.
CurEDUTx The data file that holds the test results of the most recently tested
analyzer regardless of model or serial number. The variable
x can
represent any character 0 to 9.