Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
1-22 Service Guide
The detector performance tests consists of an absolute power accuracy test and a frequency
response test as described in the HP/Agilent 86200B/86201B RF Detectors User's and
Service Guide. These tests are only for testing these two specific detector models.
Archiving of data is not done for these tests.
Additional Information
The program gives full step-by-step instructions along with connection diagrams. For
clarity, some of the required adapters may not be depicted in the connection diagrams.
Generally, solid connection lines between devices indicate a cable connection, where as a
dotted line indicates a direct connection (adapters may be required). All required
equipment should be connected and switched on before proceeding with any test, since the
program will occasionally need to write to a device before it is shown in the connection
diagram. If the device is not connected, the program will time-out before the connection
diagram is drawn. The time-out limit is set to 30 seconds for all devices. If the program
should stop for any reason, the computer may be reset and the program run again with
little or no loss of data (if a
SCRATCH C is not performed).