8712ET/ES and 8714ET/ES Adjustments
RF Network Analyzers External Detector Gain Correction, Adjustment #103
Service Guide 2-13
Figure 2-3 External Detector Connector, Front View
Warmup Time
Allow the analyzer to warm up for one hour.
Estimated Adjustment Time
This adjustment takes approximately two minutes to perform.
Procedure
1. Press
.
2. This is a fully automated adjustment handled by the internal firmware. Follow the
instructions on the analyzer's display. If you are using the voltage reference source, note
that no connection is needed other than the 12 pin connection cable.
If the analyzer fails the adjustment test, resulting in a warning message, check the
connections, voltages, and switch positions, then rerun the test.
3. If you are not going to make any more adjustments, press
.
If you are going to make more adjustments, save the correction constant data that you
have generated so far.
a. Insert a formatted disk into the internal disk drive.
b. Press .
This creates a file (or writes over an existing file) where the correction constants are
stored.
PRESET SYSTEM OPTIONS
Service
Test and Adjustments
Select Adjustment
103 ENTER
Execute Test
SYSTEM OPTIONS
Service
Update Corr Const
Store CC to EPROM
SYSTEM OPTIONS
Service
Update Corr Const
Store CC to Disk