8712ET/ES and 8714ET/ES Service Related Menus
RF Network Analyzers Update Corr Const Menu
Service Guide 5-11
Update Corr Const Menu
For a complete description of when and how to use these keys, refer to “Storing and
Recalling Correction Constants (CC)” on page 8-2. The following is a brief description.
: Loads permanent copy of correction constant data from disk to
EPROM. This key combines the functions of and
: Displays a banner which describes the keys of this menu.
: This copies the contents of the correction constant file from the
internal floppy disk drive into RAM. You cannot select an alternate disk drive. It allows
you to copy a partially modified set of data into RAM and then continue with the rest of the
adjustments.
: This will copy the current contents of the analyzer's CCs data
that is contained in RAM to permanent storage in the EPROM.
: At power-up, correction constants are copied from the flash EPROM
to a RAM buffer. Any changes to the correction constants are made to the buffer only.
Pressing this key creates a disk file that contains the correction constants from the buffer.
This file is always written to the internal disk drive, regardless of the disk drive selected
using the save/recall functions. The file will write over any existing file with the same
name.